Chapter 2

Configuration and Installation

© National Instruments Corporation

2-7

SC-2040 User Manual

To close a switch, push down the number side. Closing switch A selects a gain of 500. Closing
switch B selects a gain of 200. Closing switch C selects a gain of 100. Closing switch D selects
a gain of 10. Opening all the switches selects a gain of 1. You can select other gains by closing
more than one switch, as shown in Table 2-7.

Table 2-7. Switch Settings for Gain Selection

Desired Gain

Close Switches

1

None

10

D

100

C

200

B

300

B and C

500

A

600

A and C

700

A and B

800

A, B, and C

Input Mode Selection

Jumpers W2 through W9 select the input mode for each channel on the SC-2040. Position A-B
leaves the negative input of the instrumentation amplifier connected only to the front connector.
This is the factory-default setting. Position B-C connects the negative input of the
instrumentation amplifier to the board analog ground through a 100 k

resistor. This setting is

useful for keeping floating, or nonground-referenced, sources from saturating the instrumentation
amplifier. Table 2-8 shows the input mode jumper and channel selections.

Table 2-8. Input Mode Selection Jumpers

Channel

Use Jumper

0

W2

1

W3

2

W4

3

W5

4

W6

5

W7

6

W8

7

W9