Signal Connections

Chapter 3

SC-2040 User Manual

3-8

© National Instruments Corporation

Figure 3-4 illustrates how to connect a floating signal.

IN-

IN+

SC-2040

V

in

CH+

CH-

A

100 k

A B C

Figure 3-4. Floating Signal Connection

AC-Coupled Signals

Note: For AC-coupled signals, you must reference the analog ground of your

instrumentation to the DAQ board. The SC-2040 connector J5 provides the analog
reference, connecting directly to the MIO-16E board as shown in Figure 3-2.

For AC-coupled signals, set jumpers W2 through W9 to position B-C with an external resistor
from the positive input channel connected to its negative ground. Doing this creates the DC path
for the positive input bias current. Typical resistor values range from 100 k

to 10 M

. This

solution, although necessary in this case, lowers the input impedance of the channel and
introduces an additional offset voltage proportional to the product of the input bias current and
the resistor value used. The inputs of the SC-2040 have a typical bias current of about

±

100 pA.

When you use a 1 M

resistor, the result is

±

100

µ

V of offset, which is insignificant in most

applications. However, if you use larger valued bias resistors, significant input offset may result.
Lower valued bias resistors will increase loading of the source, possibly resulting in gain error.