NI 9217 Operating Instructions and Specifications

14

ni.com

Conversion time

High-resolution mode................. 200 ms per channel,

800 ms total for all channels

High-speed mode........................ 2.5 ms per channel,

10 ms total for all channels

Temperature accuracy (including noise)

1

, 4-wire mode

Temperature accuracy (including noise)

1

, 3-wire mode

1

For high-speed mode, add a 0.1 °C error.

Measured Value

Typical (25 °C)

Maximum

(–40 to 70 °C)

–200 to 150 °C

0.15 °C

0.35 °C

150 to 850 °C

0.20 °C

1.0 °C

Measured Value

Typical (25 °C)

Maximum

(–40 to 70 °C)

–200 to 150 °C

0.20 °C

0.50 °C

150 to 850 °C

0.30 °C

1.0 °C