Safety, Maximum voltage, Isolation voltages – National Instruments NI 9217 User Manual

Page 19

Advertising
background image

© National Instruments Corp.

19

NI 9217 Operating Instructions and Specifications

Safety

Maximum Voltage

1

Connect only voltages that are within the following limits.

All terminals-to-COM.........................±30 V

Isolation Voltages

Channel-to-channel...........................None

Channel-to-earth ground

Continuous ................................. 250 V

rms

, Measurement

Category II

Withstand.................................... 2,300 V

rms

, verified by a 5 s

dielectric withstand test

Measurement Category II is for measurements performed on
circuits directly connected to the electrical distribution system.
This category refers to local-level electrical distribution, such as
that provided by a standard wall outlet, for example, 115 V for U.S.
or 230 V for Europe.

1

The maximum voltage that can be applied between any channel

or V

sup

terminal and

a COM terminal without damaging the module or other devices.