Shock and vibration (pxi-6509 only), Safety, Electromagnetic compatibility – National Instruments NI 6509 User Manual

Page 22

Advertising
background image

NI 6509 User Guide and Specifications

22

ni.com

Shock and Vibration (PXI-6509 Only)

Operational shock ...................................30 g peak, half-sine, 11 ms pulse

(tested in accordance with
IEC-60068-2-27; test profile
developed in accordance with
MIL-PRF-28800F)

Random vibration

5 to 500 Hz, 0.3 grms .....................Operating

5 to 500 Hz, 2.4 grms .....................Nonoperating

Random vibration is tested in accordance with IEC-60068-2-64. The
nonoperating test profile exceeds the requirements of MIL-PRF-28800F,
Class 3. Random vibration is tested in accordance with IEC-60068-2-64.
The nonoperating test profile exceeds the requirements of
MIL-PRF-28800F, Class 3.

Safety

This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:

IEC 61010-1, EN 61010-1

UL 61010-1, CSA 61010-1

Note

For UL and other safety certifications, refer to the product label or visit

ni.com/

certification

, search by model number or product line, and click the appropriate link

in the Certification column.

Electromagnetic Compatibility

This product is designed to meet the requirements of the following
standards of EMC for electrical equipment for measurement, control,
and laboratory use:

EN 61326 EMC requirements; Minimum Immunity

EN 55011 Emissions; Group 1, Class A

CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A

Note

For EMC compliance, operate this device according to product documentation.

Advertising