Power requirements, Bus interface – National Instruments NI USB-9229 User Manual

Page 16

Advertising
background image

NI USB-9229/9239 User Guide and Specifications

16

ni.com

NI USB-9229...................................0.045°/kHz + 360°· f

in

/M

clk

1

NI USB-9239...................................0.075°/kHz + 360°· f

in

/M

clk

1

Phase nonlinearity (f

s

= 50 kS/s).............0.11° max

Input delay

NI USB-9229...................................38.4/f

s

+ 2.6

μ

s

NI USB-9239...................................38.4/f

s

+ 3

μ

s

Passband

Frequency ........................................0.453 · f

s

Flatness (f

s

= 50 kS/s) ......................±100 mdB max

Stopband

Frequency ........................................0.547 · f

s

Rejection..........................................100 dB

Alias-free bandwidth ..............................0.453 · f

s

–3 dB prefilter bandwidth
(f

s

= 50 kS/s) ...........................................24.56 kHz

CMRR (f

in

= 60 Hz)

NI USB-9229...................................116 dB

NI USB-9239...................................126 dB

SFDR (1 kHz, –60 dBFS).......................–128 dBFS

Total Harmonic Distortion (THD)

1 kHz, –1 dBFS ...............................–99 dB

1 kHz, –20 dBFS .............................–105 dB

Power Requirements

Current consumption from USB.............500 mA, max

Suspend mode..................................2.5 mA, max

Bus Interface

USB specification ...................................USB 2.0 Hi-Speed

1

M

clk

is the master timebase.

Advertising
This manual is related to the following products: