Signal descriptions, Table 5. signal descriptions – National Instruments 6008 User Manual

Page 11

Advertising
background image

© National Instruments Corporation

11

NI USB-6008/6009 User Guide and Specifications

Signal Descriptions

Table 5 describes the signals available on the I/O connectors.

Table 5. Signal Descriptions

Signal Name

Reference

Direction

Description

GND

Ground—The reference point for the
single-ended AI measurements, bias
current return point for differential mode
measurements, AO voltages, digital
signals at the I/O connector,

+5 VDC

supply, and the

+2.5 VDC reference.

AI <0..7>

Varies

Input

Analog Input Channels 0 to 7—For
single-ended measurements, each signal is
an analog input voltage channel. For
differential measurements, AI 0 and AI 4
are the positive and negative inputs of
differential analog input channel 0.
The following signal pairs also form
differential input channels:
<AI 1, AI 5>, <AI 2, AI 6>, and
<AI 3, AI 7>.

AO 0

GND

Output

Analog Channel 0 Output—Supplies the
voltage output of AO channel 0.

AO 1

GND

Output

Analog Channel 1 Output—Supplies the
voltage output of AO channel 1.

P1.<0..3>
P0.<0..7>

GND

Input or
Output

Digital I/O Signals—You can
individually configure each signal as an
input or output.

+2.5 V

GND

Output

+2.5 V External Reference—Provides a
reference for wrap-back testing.

+5 V

GND

Output

+5 V Power Source—Provides +5 V
power up to 200 mA.

PFI 0

GND

Input

PFI 0—This pin is configurable as either a
digital trigger or an event counter input.