NI USB-6008/6009 User Guide and Specifications

Connecting Reference Single-Ended Voltage Signals

To connect reference single-ended voltage signals (RSE) to the
NI USB-6008/6009, connect the positive voltage signal to the desired
AI terminal, and the ground signal to a GND terminal.

When no signals are connected to the analog input terminal, the internal
resistor divider may cause the terminal to float to approximately 1.4 V
when the analog input terminal is configured as RSE. This behavior is
normal and does not affect the measurement when a signal is connected.

Figure 10. Connecting a Reference Single-Ended Voltage Signal

Digital Trigger

When an AI task is defined, you can configure PFI 0 as a digital trigger
input. When the digital trigger is enabled, the AI task waits for a rising or
falling edge on PFI 0 before starting the acquisition. To use ai/Start Trigger
with a digital source, specify PFI 0 as the source and select rising or falling
edge.

AI

Voltage

Source

NI USB-6008/6009

GND