National Instruments NI 9236 User Manual

Page 23

Advertising
background image

© National Instruments Corp.

23

NI 9235/9236

Accuracy, NI 9236

Stability, NI 9236

Gain drift.............................. 6 ppm/°C
Offset drift............................ 1.7

μV/V/°C

Measurement Conditions

Percent of

Reading

*

(Gain Error)

Percent of Range

†,

(Offset Error)

30 days

after cal.

(±5 °C)

1 year

after cal.

(±5 °C)

Calibrated typ (25 °C, ±5 °C)

0.02%

0.08%

0.14%

Calibrated max (–40 to 70 °C)

0.07%

0.16%

0.39%

Uncalibrated typ (25 °C, ±5 °C)

0.15%

0.79%

Uncalibrated max (–40 to 70 °C)

0.53%

1.67%

*

Exclusive of lead wire desensitization error.

Range equals 29.4 mV/V.

Calibrated errors represent offset stability following unstrained

measurement. Errors include the effect of completion resistor tolerance
and drift.