Removing the scxi-119 x from max -14 – National Instruments 1192 User Manual

Page 22

Advertising
background image

Chapter 1

Installing and Configuring the SCXI-119X

SCXI-1190/1191/1192 User Manual

1-14

ni.com

1.

Turn off the SCXI chassis power.

2.

Remove all signals connected to the SCXI-119X.

Caution

Do not remove the SCXI-119X module from a chassis that is turned on because

doing so could damage the module.

3.

Rotate the thumbscrews securing the SCXI-119X to the chassis
counter-clockwise until they are loose, but do not completely remove
the thumbscrews.

4.

Remove the SCXI-119X by pulling steadily on both thumbscrews until
the module slides completely out.

Removing the SCXI-119X from Measurement & Automation Explorer

To remove a module from MAX, complete the following steps:

1.

Double-click the Measurement & Automation Explorer icon on
your PC desktop.

2.

Expand Devices and Interfaces. Locate the chassis in the list of
installed devices.

3.

Expand Chassis.

4.

Right-click on the module or chassis you want to delete and select
Delete.

5.

You are presented a confirmation window. Click Yes to continue
deleting the module or chassis or click No to cancel this action.

Note

Deleting an SCXI chassis deletes all modules in the chassis. All configuration

information for these modules is also lost.

Your SCXI chassis and/or SCXI module(s) should now be removed from
the list of installed devices in MAX.