Connecting signals, Connecting signals -5 – National Instruments 1192 User Manual

Page 27

Advertising
background image

Chapter 2

Using the SCXI-119X

© National Instruments Corporation

2-5

SCXI-1190/1191/1192 User Manual

Connecting Signals

The SCXI-1190 has push-on SMB connectors, and the SCXI-1191/1192
have screw-on SMA connectors as shown in Figure 2-3.

Figure 2-3. SCXI-1190 SMB Connectors, SCXI-1191/1192 SMA Connectors

Note

The power-up and reset state of the different multiplexer banks on the

SCXI-1190/1191 are as follows: banks A and C, channel 0 is connected to the COM
terminal; banks B and D, channel 3 is connected is connected to the COM terminal. The
reset mode of the SCXI-1192 connects NC to COM for all relays. Because it uses latching
relays, the SCXI-1192 does not have a specific power-up state. Its relays will retain their
state if they lose power.