Figure 2-2 illustrates the key functional components of the SCXI-1192
relay module.

Figure 2-2. SCXI-1192 Module Block Diagram

RF Switching Precautions

The SCXI-119X are

reflective switches

, meaning that any channels not

switched to the common channel are unterminated, and any signal on an
unterminated channel is reflected to its source. For most low-power
switching applications, this reflection is not a problem. However, operation
with an unterminated output can damage some high-power RF sources.
Consult your RF source documentation for more information about
connecting to unterminated channels.

Warning

Do not exceed the channel-to-ground rating of 24 V

rms

or DC on the SCXI-1190,

30 V

rms

or DC for the SCXI-1191, or 30 V

rms

or DC for the SCXI-1192. Any connections

that exceed the maximum voltage for the SCXI-119X can result in an electrical shock
hazard and damage to the switch module and any or all of the modules connected to the

SCXIbus