Wiring the NI USB-9219 Device

Caution

A high voltage screw terminal backshell must be installed when using hazardous

voltages (>42.4 V

pk

, 60 VDC).

Note

Table 2 illustrates the accessories available from

ni.com

for use with the

NI USB-9219.

The NI USB-9219 has four 6-terminal connectors that provide connections
for four analog input channels. Connect the positive signal of the signal
source to the positive input signal terminal (HI) and the negative signal of
the signal source to the negative input signal terminal (LO). Use the
excitation terminals if your sensor requires a separate excitation
connection. Refer to Table 3 for the signal names and Table 4 for the
terminal assignments for each mode. Refer to the

NI USB-9219 Circuitry

section for information about connections in each mode.

Table 2. Accessories

Accessory

Accessory Description

NI 9972

Strain relief for 6-position spring terminal

NI 9973

6-position connectors

Table 3. Signal Names

Module

Terminal

Signal Name

Signal Description

1

T+

TEDS Data

2

T–

TEDS COM

3

EX+/HI

*

Positive excitation or input
signal

4

HI

Positive input signal

5

EX–/LO

*

Negative excitation or input
signal

6

LO

Negative input signal

*

Depending on the mode, terminals 3 and 5 are either the excitation signals or the input signals.

1
2
3

4
5
6

1
2
3

4
5
6

1
2
3

4
5
6

1
2
3

4
5
6

Ch 0

Ch 1

Ch 2

Ch 3