Wiring teds channels, Grounding and shielding considerations – National Instruments NI USB-9219 User Manual

Page 12

Advertising
background image

NI USB-9219 User Guide and Specifications

12

ni.com

Wiring TEDS Channels

The NI USB-9219 supports only Class II TEDS sensors. Connect the two
TEDS lines to TEDS Data (T+) and TEDS COM (T–) and ensure that
neither T+ nor T– is tied in common to any of the signal inputs (terminals
3 through 6) on the NI USB-9219. Visit

ni.com/info

and enter the info

code

rdteds

for information about TEDS sensors.

Grounding and Shielding Considerations

You can connect ground-referenced or floating signal sources to the
NI USB-9219. If you make a floating connection between the signal source
and the NI USB-9219, make sure the voltages on the positive and negative
connections are within the channel-to-earth working voltage range to
ensure proper operation of the NI USB-9219. Refer to the

Specifications

section for more information about operating voltages and overvoltage
protection.

Note

For best signal quality, National Instruments recommends using shielded cables and

twisted-pair wiring whenever possible.

Figures 8 and 9 illustrate connecting grounded and floating signal sources
to the NI USB-9219 in Voltage mode.

Figure 8. Connecting a Grounded Signal Source to the NI USB-9219

Figure 9. Connecting a Floating Signal Source to the NI USB-9219

+

Twisted-Pair

Wiring

V

SIG

Signal Source

Shielding

HI

LO

NI USB-9219

+

Twisted-Pair

Wiring

V

SIG

V

SIG

Ground

Reference

Signal Source

Shielding

HI

LO

NI USB- 9219