Vii. microprocessor control (continued) – Traulsen TU044HT User Manual

Page 16

Advertising
background image

VII. p - ADJUSTING THE ROOM TEMPERATURE OFFSET:

The room temperature offset parameter allows a service technician or end user the ability to have the display show a tem-

perature that is within three degrees of the actual temperature being read by the cabinet air sensor. This allows for continuity

of reading between different temperature reading devices. (i.e.: thermistor vs. thermocouple vs. handheld thermometer)

This parameter is preset at the factory to “-2.5

0

F”.

Follow the instructions to enter the customer access code on page 8. When the control displays

Thermostat Set High, press the down arrow key until the control display reads

Room Temperature Offset. Press the set button . Use the arrow keys to adjust the offset to your

desired setting. When the display shows the offset you want press the set button . The display will then read

Room Temperature Offset. You can use the up or down arrow keys to scroll to the next param-

eter or press the alarm cancel button to

exit .

-15-

VII. MICROPROCESSOR CONTROL (continued)

SET

SET

)

)

)

VII. o - ADJUSTING THE DRAWER PERIMETER HEATERS:

This parameter allows the customer to turn ON and OFF the anti-condensate door/drawer perimeter heaters. This param-

eter is set to the highest setting (100) at the factory so that the door/drawer heaters stay on continuously. If you choose to

have the door/drawer heaters cycle on and off, lower this parameter to approximately 30. If condensation forms around

the door/drawers, increase the parameter until condensation stops. The exact setting will vary depending on ambient
conditions.

Follow the instructions to enter the customer access code on page 8. When the control displays

Thermostat Set High, press the down arrow key until the control display reads

DewPoint Compensation Factor. Press the set button . Press the arrow keys to adjust the factor to

your desired setting. When the display shows the factor you want press the set button . The display will then

read DewPoint Compensation Factor. You can use the up or down arrow keys to scroll to the

next parameter or press the alarm cancel button to exit .