6712SR Refrigerated Sampler
Section 4 Standard Programming

4-32

17

b

FLOW BETWEEN

SAMPLE EVENTS:

2 PULSES (1-9999)

If the module is being used for level only, enter the number of pulses
between sample events.

18

SEQUENTIAL

BOTTLES/SAMPLE

SAMPLES/BOTTLE

Select BOTTLES/SAMPLE.

19

2 BOTTLES PER

SAMPLE EVENT (1-24)

Enter 2.

20

RUN CONTINUOUSLY?

YES

NO

Select NO.

21

SAMPLE VOLUME:

250 ml (10-1000)

Enter 250.

22

NO DELAY TO START

DELAYED START

CLOCK TIME

WAIT FOR PHONE CALL

Select CLOCK TIME.

23

START FLOW COUNT AT:

06:00

Enter 6, then enter 0.

This is the time at which the sampler will start the pacing countdown.

24

SELECT DAYS:

SU MO TU WE TH FR SA

DONE

Select MO, WE, and FR.
Use the arrow keys and

to select MO, WE, and FR. When the correct days

are blinking, select DONE and press

.

25

PROGRAMMING COMPLETE

RUN THIS PROGRAM

NOW?

YES NO

Select NO.

Table 4-4 Programming Example With 730 Module Installed (Continued)