Table 5-2 storm water runoff sampling (continued) – Teledyne 6712SR User Manual

Page 98

Advertising
background image

6712SR Refrigerated Sampler
Section 5 Extended Programming

5-10

48

PACED BY:

FLOW PULSES

FLOW MODULE VOLUME

Select FLOW MODULE VOLUME.

49

FLOW BETWEEN

SAMPLE EVENTS:

2500 gal

(0.001-99999)

Enter 2500.

50

SAMPLE AT START?

YES NO

Select YES. This will cause the first sample to be taken 20 minutes (the
delay to start of sampling specified in Step 67) after the enable for Part B.

51

1 BOTTLES PER

SAMPLE EVENT (1- 18)

Enter 1.

52

SWITCH BOTTLES ON:

NUMBER OF SAMPLES

TIME

Select NUMBER OF SAMPLES.

53

SWITCH BOTTLES EVERY

1 SAMPLES (1- 50)

Enter 1.

54

RUN CONTINUOUSLY?

YES

NO

Select NO.

55

SAMPLE VOLUME:

200 ml (10-1000)

Enter 500.

56

ENABLE:

RAIN LEVEL FLOW

‘A’ DONE NONE

Select RAIN.

57

ENABLE: RAIN

AND

OR

DONE

Select AND.

58

ENABLE: RAIN AND

LEVEL FLOW ‘A’ DONE

DONE

Select LEVEL.

59

“RAIN” SET POINT:

0.15 INCHES PER

__HOURS, __MINUTES

Enter 0.15 for inches, then enter 30 minutes.

60

ENABLED WHEN:

ABOVE SET POINT

BELOW SET POINT

Select ABOVE SET POINT.

Table 5-2 Storm Water Runoff Sampling (Continued)