VXI TECHNOLOGY SVM2608 User Manual

Page 21

Advertising
background image

www.vxitech.com

SVM2608 Introduction

21

O

PTION

1

-

SVM2608-01

N

UMBER OF

C

HANNELS

2

VME

BUS

I

NTERFACE

Address Mode
Data Transfer Mode

A32
D16 or D32

S

AMPLING

R

ATE

Range
Resolution
Accuracy

7.15 Samples/s to 20.0 MSamples/s
8.333 ns
1%

M

EMORY

1 MSamples per channel

V

OLTAGE

M

EASUREMENTS

Range
Resolution
Accuracy

±0.5 V, ±1.0 V, ±2.0 V, ±5.0 V, ±10.0 V, ±20.0 V
Range/2

11

1%

I

NPUT

I

MPEDANCE

1 MΩ or 50 Ω (software selectable)

I

NPUT

M

ODE

ac or dc (software selectable)

I

NPUT

F

ILTER

5 MHz (-3 dB)

T

RIGGER

L

EVELS

Internal
Range

Trigger level resolution

Level

accuracy

External
Range

Trigger level resolution

Level

accuracy


trigger level determined by selected voltage range (see Voltage Measurement above)
Full scale/2

11

1%

-10 V to +10 V
4.88 mV
1%

D

ELAYED

T

RIGGER

Range
Resolution

0 samples to (2

24

-1) samples

1 sample

T

RIGGER

A

CCURACY

1%

E

XTERNAL

T

RIGGER

A

CCURACY

1%

T

IMEOUT

Range
Resolution

10 µs to 227 hrs
10 µs to 100 s

W

ARM

-

UP

T

IME

10 min

MTBF

80,000 hrs (based on 20% Ground Mobile, 80% Ground Fixed environment, T ≥ 52°C)






Advertising