Remote Operation
SCPI Commands for Digital Interfaces

134

Operating Manual for XDC Series Power Supply

Clear

Protection

Event

To clear a protection mechanism that has tripped:
OUTP:PROT:CLE

This will clear all protection mechanisms and re-enable the output. If the condition
that caused the alarm still exists, the protection will be allowed to trip again.

View Power

Supply

Output

The following 3 commands query the voltage, current, or power being supplied at
the output terminals.

To measure the voltage output:
MEAS:VOLT?

To measure the current output:
MEAS:CURR?

To measure the power output:
MEAS:POW?

Configure

Auxiliary

Status Lines

To set up Auxiliary line A:
OUTP:AUXA:SOUR <aux-line-mnemonic>

The choices for <aux-line-mnemonic> include:

•

NONE

•

ON: Output On

•

OFF: Output Off

•

OVOL: Over-Voltage Condition

•

UVOL: Under-Voltage Condition

•

OCUR: Over-Current Condition

•

UCUR: Under-Current Condition

•

OPOW: Over-Power Condition

•

UPOW: Under-Power Condition

•

ACOF: AC Power had been turned off or failed

•

OTEM: Over-Temperature Condition

•

HTEM: High-Temperature Condition

•

SPR: Sense Protection Tripped

•

FOLD: Fold Protection Tripped

•

CC: Constant Current Mode

•

CV: Constant Voltage Mode

•

CP: Constant Power Mode

•

UNR: Output Unregulated