9

Output data

Name of item

Header

Output data (ASCII code)

Pattern

Counting mode

Weight mode

Control data

{0

Fix to 16

2 Bytes fi xed

Control data

˜0

Fix to 1

1 Bytes fi xed

Model

MO

“KP-601”

8 Bytes fi xed

Displayed weight

Wg

x.xxx ~ xxx.xxx

5–7 bytes

Tare weight

Pt

x.xxx ~ xxx.xxx

5–7 bytes

Pieces

Pi

xxxx

1–5 bytes

Note:
• The data are divided with commas (,) for each data.
• The terminator (end of the data) is CR (ASCII code 0DH), LF (ASCII code 0AH).
• The control data for the items 1–2 are for expansion. They are not used presently so the

receiving side can ignore them.

• Because of its specifi cations, the scale will output an irrelevant signal when it’s turned off. Make

sure you ignore this signal and do not import it.

• The weight data are shown only in grams.

[Output data]

K

W

Displayed Weight

Tare Weight

Pieces

g

P t

P i

P

6 0 1