Programming auto power off – Tanita KP-601 User Manual

Page 8

Advertising
background image

6

Press ON/OFF to

turn off.

* When the registered weight is insuffi cient, “PC-25” is displayed. In that case, increase the

number of specimens to 25 pieces. When the stable mark is displayed, press CAL. If the register
is still insuffi cient, the display changes to show the number of specimens to be increased.
Therefore, increase the number of specimens, and press CAL again. This operation is repeated
until the registered weight becomes 1 g or more.

will be displayed when you press CAL if

the registered weight is too low to measure. Numeric numbers change as follows. (Press TARE
to increase the number of specimens)
10 → 25 → 50 → 100 → 150 → 200 → 250 → 500

11

PROGRAMMING AUTO POWER OFF

Default setting is 10 minutes
The auto power off time can be programmed between 0 (= no auto power off) to 60 minutes by
10 minutes increments according to preference.

Press CAL to display

.

When the power

is off

Press MODE to

confi rm

While pressing

MODE press

ON/OFF.

50

60

20

10

0

40

30

Wen you have made

your selection, press

MODE to confi rm.

The numeric values

are displayed.

Setting is fi nished.

Press TARE to cycle

through the time

options.

* If the power turns off before

appears, the setting is not saved.

1

2

3

4

5

6

7

8

* This scale might have an error within 1% margin when the specimen is

too light.