8 model 05100ps-430-601 general description, Introduction – American Magnetics 05100PS-430-601 High Stability Integrated Power Supply System User Manual

Page 21

Advertising
background image

Rev. 5

3

Introduction

General Description

more detail in section A.6.2 on page 157 of the Appendix. Contact AMI for

more information.

In addition to low LHe level, this input to the Power Supply System can be

used with other instrumentation as well. Other uses for this input include

faults from a cryocooler, temperature instrumentation, etc.

1.1.8

Model 05100PS-430-601 General Description

A Model 430 Power Supply Programmer and AMI Model 08150PS 1200

Watt unipolar voltage and current stabilized DC Power Supply are

configured with a Model 601 Energy Absorber to make up the +100 A, ±5

Vdc bipolar system designated as 05100PS-430-601. The power supply is

remotely controlled by the Model 430 Power Supply Programmer.

As a unipolar power supply, the Model 08150PS can only source

1

(not sink)

power. However, when the power supply is used in conjunction with the

AMI Model 601 Energy Absorber and controlled by an AMI Model 430

Power Supply Programmer, the result is the bipolar Model 05100PS-430-

601 integrated power supply system that is ideal for driving inductive

loads such as large magnets or motors.

The power supply is controlled by a ±10 Vdc remote analog signal supplied

by the Model 430 Programmer and applied to the power supply analog

input. Programming and control of the current loop (composed of the

magnet, power supply, Model 601 Energy Absorber, and Model 430

Programmer zero flux device), is provided by a Model 430 ramp-generated

current reference with parameters as set by the user in the Model 430. The

Model 430 compares the measured current (via the zero flux device) with

the current reference to provide precise closed-loop control of the actual

current.

The power supply is operated in voltage-voltage

2

programming mode, with

the Model 430 Programmer output scaled to operate the power supply over

its available voltage output range. The Programmer signal will continually

adjust the power supply output voltage to automatically regulate the

power supply current; precise linear power supply current control will

result as long as the system voltage and current demand do not exceed the

power supply rating or load limiting parameters.

1. The power supply is operating as a source if the current direction and voltage polar-

ity are the same (i.e., the situation that would exist when supplying a resistive load).
If the voltage polarity and current direction are opposite, the supply is operating as
a sink.

2. Voltage reference controlling voltage output.

Advertising