Gas sampling and gas conditioning technology

4-2.1.2-ME

The PSS-10/1 gas conditioning system is available with either 230V/50 Hz or with 115V/60Hz. A 6,3A
fuse is used on all models as fuse protection. The fuse is located in the power socket of the PSS
enclosure. In the event that a temperature controller is used in conjunction with heated sample lines,
the overload protection level is increased to 10A.


The electrical connection is carried out by means of a cold appliance plug and 2m of cable located on
the left-hand side of the case. The cold appliance plug socket is equipped with two-polled main
switch.

Option ‘Heated Sample Lines’:


A connector box, as shown in diagram 3, is available for the mains connection when using a heated
sample line with a PT-100 temperature sensor, or for the connection of other heated components
such as a heated sample probe or a heated filter.


the maximum length of the heated sample line which can be used is calculated as follows:


max. connection power [W] - power draw of heated components (i e. sample probes) [W]

L(m) =

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

power draw of heated sample line [W/m]