Trouble shooting – M&C TechGroup PSS-10_1 Operator's manual User Manual

Page 18

Advertising
background image

18

Gas sampling and gas conditioning technology

4-2.1.2-ME

16

TROUBLE SHOOTING

The following table aims to point out possible operational problems and offer solutions to such
problems (not applicable during the starting procedure).

Indication

Problem

Possible Cause

Check/Solution










Upper LED on
cooler does not
come on;





Middle LED on
cooler is green;
















Alarm LED on
the LA electro-
nics is red

Interruption of
gas flow;















Cooler in
operation but
gas flow
interrupted;























No voltage;







Cooler does not
function.
Cooler alarm registers
‘excess temperature’.
Cooler turns gas
measuring pump of
automatically.


Membrane pump
faulty;

Contamination of the
membrane pump;


Sample probe/hose
clogged up or lead
squashed;

Sample gas hose to
analyser clogged up or
squashed;


Optional liquid alarm
sensor:
Sensor turns
measuring pump off
automatically;











Optional flow meter:
Needle valve closed.

Check supply voltage with model type plate;
OK?
Check whether the supply voltage plug is insert
correctly and if the main switch is turned in position
‘I’
OK?
Check the fine fuse in the power socket;
OK?

Ambient temperature too high.
OK?
Free convection in case impaired  case
temperature too high;
OK?
Cooler faulty (for check see manual ECP20-1,
3-1.10ME).
OK?
Check voltage on clamps X1/11 and X1/8 ;

OK?
Remove the hoses at head of pump and check;
OK?
Clean pump if necessary;
OK?
Remove sample hose at gas inlet;
Gas flow?
Clean contaminated lead or replace;
No gas flow?
Remove sample gas hose on analyser side and
check via hose thread if sample gas flows;
No gas flow?
Clean contaminated leads or replace;
Gas flows?


Momentary overloading of the cooler due to
excessive amount of condensate;
OK?
Check hoses for condensate removal;
OK?
Check Pre-filter and if necessary replace;
OK?
Check hoses of the peristaltic pump (see manual
SR25.1, 3-7.1-ME);
OK?
Check peristaltic pump SR25.1 (see manual
SR25.1, 3-7.1-ME);
OK?
Check cooler according to instruction manual;

Adjust needle valve(s) to the desired flow.