M&C TechGroup PSS-10_1 Operator's manual User Manual
Page 22

22
Gas sampling and gas conditioning technology
4-2.1.2-ME
Sample gas IN
Bulkhead union
Back ferrule 6mm Øi
Back ferrule
10mm Øi
Sealing ring
10mm Øi
Sealing ring
6mm Øi
Insert
4mm Øo
Union nut 10mm Øi
Heated sample line
Adapter
PSS...
C o n n e c t i n g t h e h e a t e d s a m p l e l i n e w i t h s p e c i a l a d a p t e r
Place the special adapter on the PTFE hose according to the drawing seen above;
Place the insert into the PTFE hose;
Place the PTFE hose over the nipple on the bulkhead union ‘sample gas IN’ and screw the
adapter tight by hand;
Screw the adapter with a wrench (size 14) 1 1/4 rounds; for this fix the bulkhead union with a
wrench (size 15);
Place the 10mm tube of the heated sample line on the adapter and fix it by screwing the union nut
tight by hand;
Screw the union nut with a wrench (size 19) 1 1/4 rounds; for this fix the adapter with a
corresponding wrench;
Now the threaded joint is mounted tightness and can be loosen and fixed at any time.