MB23 / MB25 EN-7

Cleaning

Disconnect the Analyzer from the power supply before cleaning the Analyzer.

Make sure that no liquid enters the interior of the Analyzer.

Make sure the Analyzer is cooled down before cleaning.

Clean the Analyzer at regular intervals.

Housing surfaces and the temperature sensor may be
cleaned with a lint-free cloth slightly dampened with water
or a mild cleaning agent.

Glass surfaces may be cleaned with a commercial glass
cleaner.

Do not use solvents, harsh chemicals, ammonia or
abrasive cleaning agents.

Troubleshooting

Symptom / Display

Possible Cause

Remedy

Cannot turn on

No power to analyzer

Verify connections, fuse and voltage

Weight value blinking Sample weight less than 0.5g

Increase sample size

“Tare” shown

Pan weight needs set to zero

Press Tare

“Close Cover” shown

Cover needs to be closed before starting test

Close cover

Poor accuracy

Improper calibration
Unstable environment

Perform calibration
Move the analyzer to suitable location

Cannot calibrate

Unstable environment
Incorrect calibration masses

Move the analyzer to suitable location
Use correct calibration masses

Err 7.0

Time out

Err 8.1

Pan support has load during power on

Remove weight from pan support

Err 8.2

Pan support was removed prior to power on

Install pan support

Err 8.3

Weight on pan support exceeds capacity

Remove weight from the pan support

Err 8.4

Pan support was removed during weighing

Re-install pan support

Err 8.5

Cannot perform temperature calibration

Cool down the analyzer

Err 8.6

Current temperature is higher than setting

Cool down the analyzer

Err 9.5

Factory calibration data corrupted

Contact an Ohaus-authorized dealer

Err 10.3

Temperature sensor out of range – high

Contact an Ohaus-authorized dealer

Err 10.4

Temperature sensor out of range – low

Contact an Ohaus-authorized dealer

Err 10.5

Temperature sensor defective

Contact an Ohaus-authorized dealer

Err 10.6

Temperature sensor defective

Contact an Ohaus-authorized dealer

Err 10.7

Temperature out of control

Contact an Ohaus-authorized dealer

Err 11.0

AC power is unstable

Use a power line conditioner

Err 53

EEPROM checksum error

Contact an Ohaus-authorized dealer

Err 54

Loading of calibration parameters failed

Contact an Ohaus-authorized dealer

Service Information

If the troubleshooting section does not resolve or describe your problem, contact your authorized OHAUS service agent.
Please visit our web site: www.ohaus.com to locate the OHAUS office nearest you.

Accessories

Description

Part No.

Security Locking Cable

76288-01

Security Lock (Kensington

®

type)

470004-010

Temperature Calibration Kit

11113857

Sample Pans (50/Box)

80850086

Pads, Glass Fiber (200/Box)

80850087

Reusable Sample Pans (3/pk)

80850088

Reusable Deep Sample Pans (3/pk)

80252479

Reusable Wire Cage

80252477

Sample Pan Handler

80252476

50g Calibration Weight

51054-16

Description

Part No.

STP103 Thermal Printer, US

80251992

STP103 Thermal Printer, EU

80251993

STP103 Thermal Printer, UK

80251994

CBM910 Impact Printer, JP

80252041

CBM910 Impact Printer, US

80252042

CBM910 Impact Printer, EU

80252043

Cable, STP103

80252581

Cable, CBM910

80252571

Data collection Software

SW12

Protective glass

Temperature sensor