Introduction, Interface installation, Setup – Ohaus RS232 Interface Manual ru User Manual

Page 3

Advertising
background image

RS232 Interface

EN-1

INTRODUCTION

This Interface Kit is for use with the following Ohaus products: Scout Pro, Traveler
and Navigator.

INTERFACE INSTALLATION

Install the Interface module on the underside of the balance as shown.
Note: The appearance of your model may be different.

1

. Slide Blank Cover to rear.

2

. Remove the Blank Cover.

3

. Install Interface Module.

4

. Slide Module forward.

SETUP

A) The Ohaus Interface is preset to communicate using the following settings: 2400 baud, 7 bit, no

parity, no handshake. Determine the RS232 parameters required for the printer or computer that
is to be connected. See computer or printer documentation if assistance is required. If the
parameters do not match, it will be necessary to change either the balance settings, or the
computer / printer settings.

B) Upon installation of the interface, the balance will recognize the RS232 Interface and add

relevant items to the menu. Configure the balance to the desired RS232 and printing parameters;
refer to the balance User Manual for assistance in using the menus.

PRINT

Stable

On, Off

A-Print

Cont, On.Stbl, On.Acc*, 5sec, 15sec,
30sec, 60sec, off

End

Yes, No

*Note: Print and RS232 menu
selections may vary between
the different balance series.

RS232

RS232

On / off

Baud

600, 1200, 2400, 4800, 9600, 19200

Parity

7-even, 7-odd, 7-none, 8-none

Handshake

None, Xon-Xoff, RTS-CTS

END

Yes, No