GEOGRAPHICAL FACTOR (For Europe Only)
Press the G/N/T/
Menu button, gEO 19 is displayed. This is the current geographi-
cal adjustment value.
The geo factor includes settings from 0 to 31 and is used to compensate for
varying gravity at different geographical areas (complete geographical adjustment
settings are listed in the following table).
NOTE:
Only an authorized manufacturer’s representative or certified verification
personnel may make these changes.
Changing the geographical setting alters the calibration values.
Press the Print/
Units button until the desired geographical adjustment value is
reached. The factory default setting is gEO 19
SAVE
Press the G/N/T/
Menu button to end this block, SAVE is displayed.
Press the G/N/T/Menu button to save the geographical factor setting. The next
menu SEt uP is displayed.
or
Press the Print/
Units button to go back to the CAL menu without saving changes to
the geographical setting.
QUIT
To exit from the CAL menu, press the Print/
Units button to advance to Quit. Then
press the G/N/T/