6.2 Error Codes List
The following list describes the various error codes which can appear on the display.
LoBat
Is indicated when batteries are weak. Approximately 20 minutes of operating time remain.
Error 1
Indicates an overload condition.
Error 2
Indicates an underload condition.
Error 7
EEPROM data incorrect.
Error 14
Zero exceeds ZERO% and cannot be zeroed.
Err 21
Calibration data does not match current full scale, Grad and Cal Point settings. Settings must
be restored or the Scale must be recalibrated using the current settings.
6.3 Service Information
If the Troubleshooting section does not resolve or describe your problem, you will need to contact an authorized Ohaus
Service Agent. For Service assistance in the United States, please call Aftermarket, Ohaus Corporation toll-free at (800)
526-0659. An Ohaus Product Service Specialist will be available to help you.
6.4 Accessories
Description
Ohaus Part No.
RS232 Interface Cable/SF42 Printer
80500573
RS232 Interface Cable/PC 25 Pin
80500431
RS232 Interface Cable/PC 9 Pin
80500433
Printer
SF42
6.5 Technical Data
Materials
ABS Housing
Keypad/display overlay polyester