6.1 Input Setup

Input Setup

Input Enable
Set the input switch to

or

to enable or disable the

corresponding input. 1 to 4 inputs can be used simultaneously.

Input Range:

• +/- 1.5 volt: low range

• +/- 6 volts: high range (only available on inputs 1 and 2)

Input Type

• DC input: both AC and DC components are present in the signal.

• AC input: the DC component is filtered from the input signal.

• ICP input: polarized for an ICP sensor and DC component is filtered.

• The DC filter is a high-pass filter at Fc=0.5Hz. Common ICP sensors

are ICP microphones and accelerometers.

Input Description

This text control allows the user to identify the input. This information has no

impact on the behaviour of the acquisition system.

Unit name

This text control is used by the acquisition system for the label of the global

level, time and spectrum graphs. For instance, if a Pa unit is specified by the

user for input #1, the Y label of the Spectrum will be dB(Parms^2).

Unit dBref

The dB reference value is used by the interface for dB computation. For

acoustic measurement with a microphone, the standard dB reference value is

2E-5.