1 calibration interface – Soft dB Opus Suite Data Logger Module User Manual

Page 13

Advertising
background image

12

Sensor Sensitivity in mV/Unit.

This sensitivity is used to present a calibrated signal in the Acquisition and

Rec Viewer interfaces. The sensitivity can be manually entered (if known) or

can be determined with the calibration function (see section 6.1.1, p.12).

Import/Save Sensor Bank

The save function allows saving the complete configuration of a specific input

(range, type, input description, unit name, unit dBref and sensitivity).

The Import function allows recalling the input configuration saved in a .sen

file.

Before starting to use the data logging system, we suggest generating a .sen

file for all your sensors.

Note also that the entire set-up of the data logging instrument can be saved

in a .cfg file and recalled from one measurement session to another.

6.1.1

Calibration Interface

The input sensitivity can be manually changed in the text field or it can be calibrated using the

calibration function and a sensor calibrator. Click the

button within the Setup Interface in

order to launch the Calibration Interface.

Step 1

Adjust the calibration parameters

The defaults values are:

• Averaging time:

5 s

• Calibrator Level:

94 dB rms

2

• Frequency: 1 kHz