DVM4200

22.04.2011

©Velleman nv

15

9.2. DC Voltage

Range Resolution

Accuracy

660mV 100µV

±(0.5% of rdg + 5 digits)

6.6V 1mV

±(0.8% of rdg + 5 digits)

66V 10mV

660V 100mV

1000V 1V

±(1.0% of rdg +2 digits)

Input Impedance: >10MΩ for all ranges

9.3. AC Voltage

Range Resolution

Accuracy

660mV 100µV

±(1.0% of rdg + 5 digits)

60Hz

6.6V 1mV

40~400Hz

66V 10mV

660V 100mV

1000V 1V

±(1.5% of rdg +2 digits)

Input Impedance: >10MΩ for all ranges

Frequency Range: 40 to 400Hz

Overload protection: 1000Vdc or ac rms.

9.4. DC Current

Range Resolution

Accuracy

660µA 0.1µA

±(1.0% of rdg + 3 digits)

6.6mA 1µA

66mA 10µA

±(1.5% of rdg + 3 digits)

600mA 0.1mA

10A 10mA

±(1.8% of rdg +5 digits)

When measuring currents > 7A, max. 15s continuous measurement followed by a 15

minutes break between 2 measurements.

Overload Protection: 600mA range F0.63A/1000V– 10A range fuse F10A/1000V

9.5. AC Current

Range Resolution

Accuracy

660µA 0.1µA

±(1.5% of rdg + 5 digits)

6.6mA 1µA

66mA 10µA

±(1.8% of rdg + 8 digits)

600mA 0.1mA

10A 10mA

±(2.0% of rdg + 8 digits)

When measuring currents > 7A, max. 15s continuous measurement followed by a 15

minutes break between 2 measurements.

Overload Protection: 600mA range F0.63A/1000V– 10A range fuse F10A/1000V

9.6. Resistance

Range Resolution

Accuracy

660.0Ω 0.1Ω

±(1.2% of rdg + 2 digits)

6.600kΩ

1Ω

66.00kΩ 10Ω

660.0kΩ 100Ω

6.600MΩ 1kΩ

66.00MΩ 10kΩ

±(2.0% of rdg + 5 digits)

Note: in range 660.0Ω, short circuit test leads to determine lead resistance, than

subtract this value from the measured value.