Establishing Power to Electronics

Once the two power input wires of the shielded cable are properly connected to the terminals inside the Ex
enclosure as described above, connect the other end of the two wires to a suitable 12-24 VDC power supply
such as a battery, PLC, DCS, etc.

The digital display responds instantaneously. When power is applied, the transmitter performs several self-
diagnostic system status checks termed as “START-UP TEST” as illustrated below:

START-UP TEST

ELECTRONICS – PASS
TEMP SENSOR – PASS
BAROMETRIC SENSOR – PASS

REV. 2.14

After self diagnostic tests, the analyzer turns itself into the sampling mode. And displays oxygen contents
the sensor is exposed to, the analysis range, the ambient temperature and pressure.

0.3 %

AUTO SAMPLING

0-1% RANGE



76 F 100 KPA

Menu Navigation

The four (4) pushbuttons located on the front of the transmitter control the micro-processor functions:

Blue ENTER (select)

Yellow UP ARROW

Yellow DOWN ARROW

Green MENU (escape)