Troubleshooting, Symptom possible cause recommended action – Analytical Industries GPR-1800 IS ATEX Trace PPM Oxygen Transmitter User Manual

Page 35

Advertising
background image

Advanced Instruments, Inc

35

8. Troubleshooting

Symptom

Possible Cause

Recommended Action

Slow recovery or

At installation, defective
sensor

Air leak in sample system
connection(s)

Abnormality in zero gas

Damaged in service -
prolonged exposure to air,
electrolyte leak

Sensor nearing end of life

Replace sensor if recovery unacceptable
or O

2

reading fails to reach 10% of lowest

range

Leak test the entire sample system: Vary
the flow rate, if the O

2

reading changes

inversely with the change in flow rate
indicates an air leak - correct source of
leak

Qualify zero gas (using portable
transmitter)

Replace sensor

Replace sensor

High O

2

reading

after installing

or replacing sensor

Transmitter calibrated
before sensor stabilized
caused by:

1) Prolonged exposure to
ambient air, worse if sensor
was un-shorted

2) Air leak in sample
system connection(s)

3) Abnormality in zero gas

Allow O

2

reading to stabilize before

making the span/calibration adjustment

Continue purge with zero gas

Leak test the entire sample system
(above)

Qualify zero gas (using portable
transmitter)

High O

2

reading

Sampling

Flow rate exceeds limits

Pressurized sensor

Improper sensor selection

Correct pressure and flow rate

Remove restriction on vent line

Replace GPR/PSR sensor with XLT
sensor when CO

2

or acid gases are

present

Advertising