Adjust 4 ma with zero o – Analytical Industries GPR-1800 IS ATEX Trace PPM Oxygen Transmitter User Manual

Page 45

Advertising
background image

Advanced Instruments, Inc

45

Adjust 4 mA with Zero O

2

Access the MAIN MENU by pressing the MENU key.

Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.

Press the ENTER key to select the highlighted menu option.

The following displays appear:

MAIN MENU

AUTO SAMPLE

MANUAL SAMPLE

CALIBRATION

>>>

CALIBRATION

SPAN CALIBRATE

ZERO CALIBRATE

DEFAULT SPAN

DEFAULT ZERO

OUTPUT SPAN

OUTPUT ZERO

Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT ZERO.

Press the ENTER key to select the highlighted menu option.

The following display appears:

100.0

OUTPUT ZERO OFFSET

PRESS UP OR DOWN

TO CHANGE VALUE

ENTER TO SAVE

MENU TO RETURN

Enter the calculated adjustment value.

Note:

Once the initial adjustment is made and checked at the PLC

it may be necessary to fine tune the initial adjustment by repeating. Any additional percent error must be
added or subtracted from the initial adjustment value

000.0

OUTPUT ZERO OFFSET

PRESS UP OR DOWN

TO CHANGE VALUE

ENTER TO SAVE

MENU TO RETURN

Press the ENTER key to advance the underline cursor right or press the MENU key to advance the
underline cursor left to reach to the desired digit of the adjustment OUTPUT ZERO OFFSET value.

Press the ARROW keys to enter each the numerical value of each digit of the adjustment OUTPUT ZERO
OFFSET value.

Repeat until the complete OUTPUT ZERO OFFSET value has been entered.

Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.

The system returns to the SAMPLING mode.