Symptom possible cause recommended action – Analytical Industries GPR-1800 IS ATEX Trace PPM Oxygen Transmitter User Manual

Page 36

Advertising
background image

Advanced Instruments, Inc

36

Symptom

Possible Cause

Recommended Action

Response time slow

Air leak, dead legs, distance
of sample line, low flow rate,
volume of optional filters and
scrubbers

Leak test (above), reduce dead volume or
increase flow rate

O

2

reading doesn’t

agree to expected
O

2

values

Pressure and temperature of
the sample is different than
span gas

Abnormality in gas

Calibrate the transmitter (calibrate at
pressure and temperature of sample)

Qualify the gas (use a portable analyzer as
a second check)

Erratic O

2

reading

or

No O

2

reading

Change in sample pressure

Dirty electrical contacts in
upper section of sensor
housing

Corroded solder joints on
sensor PCB from corrosive
sample or electrolyte leakage
from sensor

Corroded spring loaded
contact in upper section of
sensor housing from liquid in
sample or electrolyte leakage
from sensor

Liquid covering sensing area

Improper sensor selection

Presence of interference
gases

Unauthorized maintenance

Sensor nearing end of life

Calibrate the transmitter (calibrate at
pressure and temperature of sample)

Clean contacts with alcohol (minimize
exposure time of MS sensor to ambient air
to extent possible)

Replace sensor and return sensor to the
factory for warranty determination

Upper section of sensor housing: Clean
contacts with water, wipe contacts with
clean paper towel and flush system and
sensor housing with dry gas

Sensor: Replace if leaking and return it to
the factory for warranty determination

Wipe with lint free towel or flow dry sample
or zero gas for 2-3 hours to flush out
condensation

Consult factory for recommendation.

Replace sensor and install scrubber

Consult factory.

Replace sensor

Erratic O

2

reading

or

Negative O

2

reading

Pressurizing the sensor by
flowing gas to the sensor with
the vent restricted or SHUT
OFF valve closed and

Zero the transmitter. If not successful
replace the sensor