Span Gas vs. Air Calibration

The analyzer can be calibrated by using ambient air (20.9% O

2

) or a certified span gas. The only advantage

of using a certified span gas for calibration is the fast recovery time to low PPM level after calibration. For
example, if the analyzer is calibrated by using 8 PPM span gas, the analyzer will recover to within 0.1 PPM
of the original reading within five minutes. On the other hand, if the analyzer is calibrated with ambient air, it
may take 1-2 hours for the reading to get below 0.5 PPM

Zero Calibration Procedure

Zero calibration should be carried out after the span calibration has been performed. Normally, zero
calibrations are performed when a new sensor is installed or changes are made in the sample system
connections.

Before performing a ZERO calibration, it is highly recommended to perform a factory default zero. This will
eliminate previous zero offset adjustment that might have been made. With factory default setting, if the zero
offset does not exceed 50% of the lowest range, this indicates that the integrity of the sensor, the analyzer
sample system and the sample line bringing in the sample gas is maintained.

Access the MAIN MENU by pressing the MENU key.

Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.

Press the ENTER key to select the highlighted menu option.

The following displays appear:

MAIN MENU

AUTO SAMPLE

MANUAL SAMPLE

CALIBRATION

CONFIG ALARMS

BYPASS ALARMS

>>>

CALIBRATION

SPAN CALIBRATE

ZERO CALIBRATE

DEFAULT SPAN

DEFAULT ZERO

OUTPUT SPAN

OUTPUT ZERO

Advance the reverse shade cursor using the ARROW keys to highlight ZERO CALIBRATE.

Press the ENTER key to select the highlighted menu option.

The following displays appear:

ZERO

CALIBRATION

ENTER TO CALIBRATE

MENUE TO ABORT

Press the ENTER key to calibrate or MENU key to abort and return to SAMPLING mode.

Allow approximately 60 seconds for the calibration process while the microprocessor determines whether
the signal output or reading has stabilized within 50% of the full scale low range. If the offset is less than
50% of the lowest range, by pressing ENTER, message PASSED CALIBRATION will appear and return to
the Sample mode. On the other hand, if the offset is above 50%, pressing ENTER, message FAILED