18,500 ft. – 16 AWG
29,500 ft. – 14 AWG




Connections – Optional Intrinsic Safety Barrier:
Non-hazardous area: Barrier terminals (1), (2)
Hazardous area: Barrier terminals (3), (4)
Direct measurement of 4-20mA: Reference (C1), connect the current measuring device at terminal (2) of the

barrier and the negative (-) terminal from the power source as shown below.
Convert 4-20mA current to voltage: Reference (C2), connect a resistor (RL) maximum value 850Ω between
terminal (2) of the barrier and the negative (-) terminal from the power source and measure the voltage across the

resistor.
Example: To convert the 4-20mA current for output to a PLC requiring 1-5V, follow the above procedure using a

250Ω resistor (RL).



















Hazardous Area Operation: