Analytical Industries GPR-2500 S Oxygen Analyzer User Manual

Page 27

Advertising
background image




27

Press the ENTER key to select the SPAN CALIBRATE option.
Note: A span gas concentration above 1000 ppm dictates the selection of the PERCENT option.
Advance the reverse shade cursor using the ARROW keys to highlight the desired GAS CONCENTRATION.
Press the ENTER key to select the highlighted menu option.

GAS CONCENTRATION

PERCENT

PPM






The following displays appear:

00.00 %

PRESS UP OR DOWN
TO CHANGE VALUE
SELECT TO SAVE
ESC TO RETURN



>>>

20.9 %

SPAN

CALIBRATION

ENTER TO CALIBRATE

MENU TO ABORT


Press the ENTER key to advance the underline cursor right to the second digit of the span value. Press the MENU

key to advance the underline cursor left to the previous digit.
Press the ARROW keys to enter the second digit of the span value.
Repeat steps 30 and 31 until the complete span value has been entered.
Allow approximately 60 seconds for the calibration process while the processor determines whether the signal

output or reading has stabilized within 60% of the full scale low range.
Both the Zero Calibrate and Span Calibrate functions result in the following displays:



PASSED

CALIBRATION




OR


FAILED

CALIBRATION




If the calibration is successful, the monitor returns to the SAMPLING mode after 30 seconds.