Menu navigation, Main menu, Range selection – Analytical Industries GPR-2500 S Oxygen Analyzer User Manual

Page 19

Advertising
background image




19

Note: The monitor display defaults to the sampling mode when 30 seconds elapses without user interface.

20.9

%

MANUAL SAMPLING

25% RANGE


24.5 C 100 KPA

Menu Navigation

The four (4) pushbuttons located on the front of the monitor operate the micro-processor:

¾ green ENTER (select)

¾ yellow UP ARROW

¾ yellow DOWN ARROW

¾ blue MENU (escape)

Main Menu

Access the MAIN MENU by pressing the MENU key:

MAIN MENU

AUTO SAMPLE

MANUAL SAMPLE
CALIBRATE

24.5 C 100 KPA

Range Selection

The GPR-2500S Ambient Oxygen Monitor is equipped with four (4) standard measuring ranges (see specification)

and provides users with a choice of sampling modes. By accessing the MAIN MENU, users may select either the

AUTO SAMPLING (ranging) or MANUAL SAMPLING (to lock on a single range) mode.
Caution: For ambient monitoring, the user must use the MANUAL SAMPLE option to lock on to the 0-25% range.

Also, when ambient monitoring for oxygen deficiency in a confined space or room do not calibrate the unit in the

atmosphere to be monitored. Use a certified span gas of a known oxygen concentration approximating 20.9%
oxygen balance nitrogen or clean compressed air.
Procedure - Auto Sampling:
Access the MAIN MENU by pressing the MENU key.
Advance the reverse shade cursor using the ARROW keys to highlight AUTO SAMPLE.

Advertising