When used in conjunction with the optional MTL 702, third party certified, intrinsic safety barriers, the design of

the GPR-2500S Ambient Oxygen Monitors meet recognized standards as intrinsically safe for operation in Class I,

II, III; Division I, II; Groups A-G hazardous areas.
Note: Locate the optional intrinsic safety barrier as close to the power source in the non-hazardous area as

possible.


Output connection:
The 4-20mA current output is obtained by connecting the current measuring device between the negative terminal

of power source and the negative terminal, marked (-), located in the junction box of the monitors. The positive

current flow is from pin 1 to pin 2 and from pin 2 to ground through the external load.
To check the signal output of the 4-20mA E/I integrated circuit connect an ammeter, as illustrated below, as the

measuring device and confirm the output is within +0.1mA of 4mA.




Power 12-36V DC

To Monitors

Terminal Strip






Caution:
To assure proper grounding, connect the 4-20mA signal output to the external device (PLC, DCS, etc.)

before attempting any zero or span adjustments.