Zero calibration – Analytical Industries GPR-2500 S Oxygen Analyzer User Manual

Page 22

Advertising
background image




22


Zero Calibration

In theory, the oxygen sensor produces no signal output when exposed to an oxygen free sample gas. However, the

monitor will generate an oxygen reading when sampling oxygen free sample gas due to:
¾ Contamination or quality of the zero gas
¾ Minor leakage in the sample line connections
¾ Residual oxygen dissolved in the sensor’s electrolyte
¾ Tolerances of the electronic components

Recommendation: Zero calibration is recommended for measurements below 1% on the 0-1% range, as it is not

practical on higher ranges as described below. This recommendation is applicable to the GPR-2500S, however, this

section is provided for users discretion.
Procedure:
Zero calibration should precede the span calibration and once performed should not have to be repeated with

subsequent span calibrations. Normally, zero calibrations are performed when a new sensor is installed or changes

are made in the sample system connections.
Refer to Span Calibration below for the detailed procedure. Differences include the displays illustrated below,

substituting a suitable zero gas for the span gas and allowing the monitor 24 hours with flowing zero gas to

determine the true zero offset (a stable reading evidenced by a horizontal trend on an external recording device) of
the system before conducting the zero calibration. Note: 24 hours is required for the sensor to consume the

oxygen that has dissolved into the electrolyte inside the sensor (while exposed to air or percentage levels of

oxygen).
Thus, for the reasons above, it is not practical to zero a percent monitor. Finding the true zero offset is not always

necessary particularly in the case of applications requiring higher level oxygen measurements because of the low
offset value, normally < 0.1 ppm, is not material to the accuracy of higher level measurements.
Note: Prematurely zeroing the monitor can cause a negative reading in both the ZERO and SAMPLE modes.
Access the MAIN MENU by pressing the MENU key.
Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATE.
Press the ENTER key to select the highlighted menu option.
Repeat to select ZERO CALIBRATE and the following displays appear:

MAIN MENU

AUTO SAMPLE
MANUAL SAMPLE

CALIBRATE


24.5 C 100 KPA



>>>

CALIBRATION

SPAN CALIBRATE

ZERO CALIBRATE

DEFAULT SPAN
DEFAULT ZERO


24.5 C 100 KPA




Advertising