Xylem handylab LF 11 User Manual
Page 19

handylab LF 11
Operation
19
ba75365e04
01/2011
AutoRead AR
(drift control)
The AutoRead function (drift control) checks the stability of the measurement
signal. The stability has a considerable effect on the reproducibility of the
measured value.
Note
The current AutoRead measurement can be terminated at any time (accept-
ing the current value) by pressing <RUN/ENTER>.
5
Wait for a stable measured value.
1
Select the required measuring mode with <M>.
2
Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).
3
Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
4
If necessary, start the next AutoRead measurement with <RUN/EN-
TER>.
5
To terminate AutoRead: Press the <AR> key.