Operation

handylab LF 11

32

ba75365e04

01/2011

4.5

Reset

Measurement

parameters

The following measured parameters (

ϰ

InI) are reset to the default condition:

Note
The calibration data gets lost when the measuring parameters are reset.
Recalibrate after performing a reset.

Resetting the

measuring

parameters

Measuring mode

ϰ

Cell constant

0.475 cm

-1

(calibrated)

0.475 cm

-1

(set up)

Temperature compensation

nLF

Reference temperature

Temperature coefficient of the lin-
ear temperature compensation

2.000 %/K

TDS factor

1.00

1

Press the <RUN/ENTER> key and hold it down.

2

Press the <CAL> key.

3

Using <V> <W>, switch between no and YES.
YES: Resetting the measuring parameters
no: Retaining settings.

4

Confirm with <RUN/ENTER>.
The measuring instrument automatically switches to the measuring
mode.