6 what to do if, What to do if, Chapter 6 w – Xylem handylab LF 11 User Manual
Page 35: 6what to do if

handylab LF 11
What to do if...
35
ba75365e04
01/2011
6
What to do if...
LoBat display
Instrument does
not react to key-
stroke
Error message
OFL
Error message
E3
Display
to
Cause
Remedy
– Batteries almost empty
– Replace batteries
)
Cause
Remedy
– Operating condition
undefined or EMC load
unallowed
– Processor reset:
Press the <CAL> and
<ON/OFF> keys at the same
time and release them again.
The software version is
displayed.
Cause
Remedy
The measured value lies outside
the measuring range
– Measuring cell not connected
– Connect measuring cell
– Cable broken
– Replace measuring cell
Cause
Remedy
– Measuring cell contaminated
– Clean cell and
replace it if necessary
– Unsuitable calibration
solution
– Check calibration solutions
Cause
Remedy
– Time-out of the interface
– Check the instrument that is
connected