C a l – Xylem handylab multi 12 User Manual
Page 50

Operation
handylab multi 12
50
ba75371e03
11/2009
Outputting the
calibration data
on the display
1
Press the <RCL> key repeatedly until CAL disp appears on
the display.
2
Press the <RUN/ENTER> key.
The data of the last calibration of all measured variables ap-
pears in the following sequence:
z pH: Slope SLO and asymmetry ASY
z Dissolved oxygen: Relative slope SLO
z Cond: Cell constant C
Information concerning the calibration procedure is output as
well.
3
Press <RUN/ENTER> to display the value of the asymmetry
(mV).
Store