3 conductivity system messages, Conductivity system messages – Xylem handylab multi 12 User Manual
Page 70

What to do if...
handylab multi 12
70
ba75371e03
11/2009
Measured value
too low
Measured value
too high
6.3
Conductivity system messages
Error message
OFL
Error message
E3
Cause
Remedy
– Insufficient flow
– Provide flow to the probe
Cause
Remedy
– High amount of dissolved
substances
– Correct solubility function
using the salinity equivalent
– Air bubbles bump on the
membrane with high velocity
– Avoid direct flow to the
membrane
– The carbon dioxide pressure is
too high (> 1 bar)
– Measuring not possible
Cause
Remedy
The measured value lies outside
the measuring range
– Measuring cell not connected
– Connect measuring cell
– Cable broken
– Replace measuring cell
Cause
Remedy
– Measuring cell contaminated
– Clean cell and replace it if
necessary
– Unsuitable calibration
solution
– Check calibration solutions