2 oxi system messages, Oxi system messages – Xylem handylab multi 12 User Manual
Page 69

handylab multi 12
What to do if...
69
ba75371e03
11/2009
6.2
Oxi system messages
Error message
0FL
Error message
E3
Error message
E7
AR flashes
continuously
Cause
Remedy
Display range exceeded
Oxygen sensor
– Not connected
– Connect the probe
– Cable broken
– Replace probe
– Depleted
– Replace probe
– Short-circuit between gold and
lead electrode
– Clean probe and replace it if
necessary
Cause
Remedy
Invalid calibration
Oxygen sensor
– Electrolyte solution depleted
– Regenerate probe
– Membrane contaminated
– Clean membrane
– Electrode system poisoned
– Regenerate probe
– Worn out
– Replace probe
– broken
– Replace probe
Cause
Remedy
Membrane damaged
– Membrane damaged
– Regenerate probe
– Membrane head not screwed
on tight enough
– Screw membrane head tight
Cause
Remedy
No stable measured value
– Membrane contaminated
– Clean membrane