2 initial commissioning, Initial commissioning – YSI CR 3200 Thermoreactor User Manual

Page 12

Advertising
background image

Commissioning

CR 3200

12

ba76120e01

08/2012

3.2

Initial commissioning

Note
The thermoreactor works at an ambient temperature of +5 °C to
+40 °C. When the thermoreactor was transported from a cold
environment to a warm environment, condensate may occur and cause
a malfunction. Wait until the thermoreactor has adjusted to the new
environmental conditions before putting it into operation again (see
also chapter 8 T

ECHNICAL

D

ATA

).

Setting up the

thermoreactor

Adjusting the mains

voltage

1

Place the thermoreactor firmly onto a heat-resistant
underground.

2

Make sure that there is enough space between the
thermoreactor and other instruments or devices that are heat-
sensitive.

3

Check whether the arrow on the housing points to the mains
voltage (115 or 230 V) given on the fuse holder (3) that is
provided by the mains.

4

If the wrong mains voltage is set, perform steps 5 to 7.

5

Pull out the fuse holder (3).

6

Turn the fuse holder (3) so that the arrow on the housing points
to the mains voltage (115 or 230 V) provided by the mains.

7

Push the fuse holder (3) in completely.

2

3

1