8 technical data, Technical data, In chapter 8 t – YSI CR 4200 Thermoreactor User Manual

Page 33: Echnical, Chapter 8 t, 8technical data

Advertising
background image

CR 4200

Technical Data

33

ba76121e01

09/2012

8

Technical Data

Reactor type

Dry temperature control device
with safety cover

Cell shafts

2 x 12 cell shafts for reaction
cells 16 ± 0.2 mm

Reaction time setting

20 min, 30 min, 60 min, 120 min
(via fixed programs)
8 freely adjustable programs:
0...180 min

Temperature setting

100 °C, 120 °C, 148 °C, 150 °C
via fixed programs
and 8 freely adjustable
programs:
Room temperature ...170 °C

Controlling accuracy

± 1 °C ± 1 Digit

Temperature stability

± 0.5 K

Overtemperature protection

190 °C ± 5 °C

Heating time (with empty
thermoblock) from 25 °C to

100 °C approx. 5 min
120 °C approx. 7 min
148 °C approx. 10 min

Temperature of the enclose at
an environmental temperature
of 25 °C

< 30 °C with a block temperature
of 148 °C

Output

unidirectional RS232 interface
with 2 banana sockets for:
 external temperature probe

or
 PC cable

Power supply

230 VAC 50 Hz ± 15 %
115 VAC 60 Hz ± 15 %
Power consumption: 560 W
Fuses 2 x 6.3 AT

Enclosure

PC ABS, recyclable, high
temperature resistant

Protective class

I according to DIN VDE 0700
part 1/11.90

Advertising